How do you determine the appropriate level of centrifugation?
Take the liquid over the column but keep the packing wet as the standard, just follow the parameters of the instruction manual. If the whole tube consists of a large amount of liquid not centrifuged down, you can repeat the centrifugation a few times, or consider that it may be caused by too high a concentration of protein, you need to dilute the sample and take 1-10 μg of protein on the sample.
Q:
Does Dissolve B contain iRT?
No. Self-adding iRT is available.Additionally Dissolve B can be replaced with solvents commonly used in the laboratory to re-solubilize peptides.
Q:
What volume of Dissolve B is used to dissolve dried peptides for LC-MS testing?
For timsTOF instruments (Bruker), dissolve the peptide to 200 ng/μL and inject 1 μL. If the initial sample size is less than 500 ng, it is recommended that 5 μL of Dissolve B be used to dissolve the sample and 4 μL be injected; For Orbitrap instruments (Thermo), dissolve peptide to 1 μg/μL of sample, e.g., for an initial sample of 10 μg, dissolve peptide in 10 μL of Dissolve B and inject 1-3 μL. If the initial sample size is less than 2 μg, it is recommended to dissolve peptide in 5 μL of Dissolve B and inject 4 μL of peptide.
Q:
Is the Adaptor Adaptor reusable?
Yes, reuse is recommended. At the end of each experiment, the Adaptor should be submerged in anhydrous ethanol and ultrasonically cleaned (in a special glass beaker), dried naturally and then stored in a special glass beaker and sealed with tinfoil until the next use.
Q:
Can a tube of lyophilized enzyme be saved for reuse after reconstitution?
Use within 3h after reconstitution, it is not recommended to keep it for reuse.