      • Q:
        What volume of Dissolve B is used to dissolve dried peptides for LC-MS testing?
        For timsTOF instruments (Bruker), dissolve the peptide to 200 ng/μL and inject 1 μL. If the initial sample size is less than 500 ng, it is recommended that 5 μL of Dissolve B be used to dissolve the sample and 4 μL be injected; For Orbitrap instruments (Thermo), dissolve peptide to 1 μg/μL of sample, e.g., for an initial sample of 10 μg, dissolve peptide in 10 μL of Dissolve B and inject 1-3 μL. If the initial sample size is less than 2 μg, it is recommended to dissolve peptide in 5 μL of Dissolve B and inject 4 μL of peptide.
      • Q:
        Is the Adaptor Adaptor reusable?
        Yes, reuse is recommended. At the end of each experiment, the Adaptor should be submerged in anhydrous ethanol and ultrasonically cleaned (in a special glass beaker), dried naturally and then stored in a special glass beaker and sealed with tinfoil until the next use.
      • Q:
        Can a tube of lyophilized enzyme be saved for reuse after reconstitution?
        Use within 3h after reconstitution, it is not recommended to keep it for reuse.
      12